Bench Meter Comes with Data Logging Software Included
 

Mi170 Specifications

Range EC


TDS



NaCL

Temp
0.00 to 29.99 µS/cm; 30.0 299.9 µS/cm; 3.00 to 29.99 mS/cm;
30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual conductivity

0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm);
150 to 1499 mg/L (ppm); 1.5 to 14.99 g/l (ppt); 15.0 to
100.0 g/l (ppt); up to 400.0 g/l actual TDS (with .080 factor)

0.0 to 400.0%

-20.0 to 120.0°C
Resolution EC
TDS
NaCL
Temp
0.01 µS/cm; 0.1 µS/cm; 1.0 µS/cm; 0.01 mS/cm; 0.1 mS/cm
0.01 mg/L; 0.1 mg/L; 1.0 mg/L; 0.01 g/l; 0.1 g/l
0.1%
0.1°C
Accuracy EC
TDS
NaCL
Temp
± 1% of reading ± (0.05 µS/cm or 1 digit)
± 1% of reading ± (0.03 mg/L or 1 digit)
± 1% of reading
± 0.4%
Calibration EC - 1 point slope calibration with 6 memorized solutions
available (84.0 µS, 1413 µS, 5.00 mS/cm, 12.88 mS/cm,
80.0 mS/cm, 111.8 mS/cm). NaCL - 1 point,
with MA9066 calibration solution
Temp Compensation Automatic or Manual, from -20.0 to 120.0°C
Temp Coefficient Selectable from 0.00 to 6.00% / °C (EC and TDS only)
Probe MA814D/1 (4 ring electrode with temperature probe included)
TDS Factor 0.40 to 0.80 (default value is 0.50)
Log on demand 50 on each range (EC, TDS, NaCL)
GLP Last EC, NaCL calibration data
PC Interface RS232 / USB Opto-isolated
Environment 0 to 50°C / 32 to 122°F; max RH 95%